Operation
Calibration
Thermo Scientific
146iQ Instruction Manual
3-23
The Pressure Calibration Screen is used to calibrate the pressure sensors on
the Flow/Pressure DMC board to zero, span or factory default values. The
Pressure Calibration screen is only visible when there is an ozonator and/or
a photometer installed in the instrument.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Pressure Calibration
The Pressure Calibration screen contains the following information:
●
Across:
●
Description:
Lists items in table.
●
Reading:
Displays reading of each pressure sensor.
●
Calibration:
Starts calibration or resets default values.
●
Down:
●
Sensor 1
3 Reading:
Under the column labeled Reading, current
reading of each pressure sensor. Sensor 1 is connected to the pump
(vacuum) side of the photometer capillary, and in the 146iQ is the
sensor on the bottom of the board. Sensor 2 is a differential
pressure sensor that is connected to the 2 sides of the ozonator
capillary. Sensor 3 is connected to the bench (pressure) side of the
photometer capillary, and in the 146iQ is the sensor on the top of
the board. The Sensor readings are only valid if their corresponding
options are installed, and can otherwise be ignored.
●
Atmospheric Sensor 1
3:
User should expose the sensor to
atmospheric pressure by disconnecting the tube attached to the
sensor. Under the column labeled Reading, the user enters the
current atmospheric pressure in mm Hg units. Under the column
Pressure Calibration
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......