Servicing
Optional Zero Mix Valve Replacement
Thermo Scientific
146iQ Instruction Manual
7-55
7.
Replace valve with same fitting orientation related to the markings on
the valve (COM, NC, NO). Reassemble in reverse order.
Note
All National Pipe Thread fittings require PTFE tape for sealing.
Remove the existing PTFE tape and apply new. Wrap PTFE tape in the
same direction of the thread spiral. Wrap around approximate two times
around depending on the thickness of the PTFE tape. Make sure PTFE
tape does not overhang the first thread (the lead in thread).
▲
Figure 7–62.
Zero Mix Valve Replacement
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......