Optional Equipment
Permeation Oven
Thermo Scientific
146iQ Instruction Manual
9-17
permeates the gas into a zero air stream (150 sccm nominally) that can
subsequently be diluted to provide the desired span gas output. When the
permeation oven is not being actively utilized, the nominal zero air stream
through the permeation oven is maintained, in order to precisely regulate
temperature, with a solenoid directing this standby flow to exhaust.
Use the following procedure to install the permeation tube.
1.
Lift oven cover up.
2.
Lift and unlatch two oven cover handles.
Figure 9–9.
Installing glass tube pt 1
Permeation Tube
Installation
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......