Servicing
Mixing Chamber Replacement
7-62
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the mixing chamber.
Equipment required:
Phillips drive, #2
Hex drive, 9/64
Adjustable wrenches, qty 2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Using two Adjustable wrenches (one to hold fitting body and one to
loosen the nut), disconnect plumbing 3X.
3.
Using 3/32 hex drive, loosen the two setscrews (detail ‘A’).
Figure 7–70.
Mixing Chamber Replacement, Plumbing
4.
Pull mixing chamber upwards.
5.
Using adjustable wrench, disconnect tee fitting as shown.
6.
Replace mixing chamber and reassemble in reverse order.
Mixing Chamber
Replacement
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......