C-Link Protocol Commands
I/O Configuration
B-30
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Table B–9.
Analog Current Output Range Values
Range
Output Range
1 0-20
mA
2 4-20
mA
0 [cannot be set to this, but may report]
Undefined
analog vin
channel
This command retrieves the analog voltage input
channel
data, both the
calculated value and the actual voltage. In the following example, the
“calculated” value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with “feature not enabled” if the I/O expansion board
is not detected.
Send:
analog
vin
1
Receive:
analog vin 1 75.325 2.796
analog vout range
channel
This command reports the analog voltage output
channel
range, where
channel
is between 1 and 6 inclusive, according to
Table B–10
. The
following example reports that analog voltage output channel 2 is set to 3
(1-10 V).
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range
channel range
This command sets analog voltage output
channel
to the range, where
channel
is between 1 and 6 inclusive, and
range
is set according to
Table B–
10
. The following example sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B–10.
Analog Voltage Output Range Values
Range
Output Range
1 0-1
V
2 0-100
mV
3 0-10
V
4 0-5
V
0 [cannot be set to this, but may report]
Undefined