Firmware Screen Descriptions
Instrument Setup Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-33
In the Main Menu, choose Instrument Setup >
Datalogging >
Reset to
Default Content
.
●
●
●
The Configure Datalogging menu is used to configure the datalogging for
the currently selected record type.
●
In the Main Menu, choose Instrument Setup >
Datalogging >
Configure Datalogging
.
DATALOGGING SETTINGS:
>LOGGING PERIOD MIN 1
MEMORY ALLOCATION % 70
DATA TREATMENT AVG
FLAG STATUS DATA OFF
CTRST SSET INSTS STAT
The Logging Period Min screen is used to select the time period in minutes
between logged data points for the selected record type (srec or lrec). The
list of choices includes: Off, 1, 5, 15, 30, and 60 minutes (default).
●
In the Main Menu, choose Instrument Setup >
Datalogging >
Configure Datalogging >
Logging Period Min
.
SET PERIOD FOR USER:
CURRENTLY: 1 MIN
SET TO: 1 MIN
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Memory Allocation Percent screen is used to select the percentage of
each record type for both srecs and lrecs. The “other” type is set to 100%
which is the specified value. Percentages between 0 and 100% are available
Configure Datalogging
Logging Period Min
Memory Allocation
Percent
RESET USER DATA AND
ERASE USER LOG FILE DATA?
RESET
CTRST SSET INSTS STAT
RESET USER DATA AND
ERASE USER LOG FILE DATA?
RESET
ARE YOU SURE YOU WANT TO?
PRESS
TO CONFIRM RESET
xx
CTRST SSET INSTS STAT