Communications Using iPort or RPComm
Using RPComm
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
D-37
&D0 Ignore DTR (Data Terminal Ready) from instrument
S0=1 Set auto answer to one ring
&K0 Disable local flow control
&W0 Store settings in profile “0”
&Y0 Use stored settings in profile “0” on power up.
7.
Disconnect your computer from the modem.
8.
Attach the male end of the 9-to-9 pin cable to the RS232 connector on
one of the instruments.
9.
Attach the null modem adapter to the female end of the 9-to-9 pin
cable.
Note
Not all null modem adapters can be used in this application because
pin connections are not standard. The null modem adapter (10-005671)
that is available from Thermo Fisher Scientific is acceptable for use with
the Sampler.
▲
10.
If your modem is equipped with a 9-pin connector, attach the cable
and null modem adapter to the modem. If your modem is equipped
with a 25-pin connector, attach a 9-to-25 pin serial adapter to the null
modem adapter and plug it into the modem.
Note
Not all 9-to-25 pin serial adapters can be used in this application
because pin connections are not standard. The 9-to-25 pin serial adapter
(51-001079) that is available from Thermo Fisher Scientific is acceptable
for use with the Sampler.
▲
11.
Attach a phone line to the modem.
12.
With the RPComm Connect List screen displayed (
Figure D–20
),
select the name of one of the connections used to verify the direct
connection above and choose the Edit Selected Connection icon. The
Connection Type screen will then be displayed (
Figure D–21
).
13.
From the Connection Type screen, ensure that the correct instrument
type is selected and select “Settings” to display the AK Protocol Setup
screen (
Figure D–22
).