Firmware Screen Descriptions
Service Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-81
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, individual voltage channels 1-6, and
individual current channels 1-6 (if the I/O expansion board option is
installed). Configuration choices include selecting range, setting
minimum/maximum values, and choosing signal to output. The Allow
Over/Under Range item scales the output to allow the operator to detect
signals that have gone outside the normal range.
●
In the Main Menu, choose Service > External I/O >
Analog Output
Configuration
.
OUTPUT CHANNELS:
>ALLOW OVER/UNDER RANGE
ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
CTRST SSET INSTS STAT
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1-6
The Allow Over/Under Range screen is used to select whether or not the
analog outputs are allowed to exceed the maximum selected value of 100
mV, 1 V, 5 V, 10 V, or 20 mA or the minimum selected value of 0 V, 0
mA, or 4 mA. By default this parameter is set to on, and 5% over and
under range is allowed for all analog output channels.
●
In the Main Menu, choose Service > External I/O > Analog Output
Configuration >
Allow Over/Under Range
.
ALLOW OVER/UNDER RANGE:
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
CTRST SSET INSTS STAT
The Analog Output Configuration menu displays a list of the analog
output configuration choices from which the user selects the parameter to
adjust for the selected output channel. Configuration choices include
Analog Output
Configuration (Select
Channel)
Allow Over/Under Range
Analog Output
Configuration (Select
Action)