Firmware Screen Descriptions
Service Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-73
The Number of Items menu selection is used to set the number of
instrument readings to display on the custom Run screen. A maximum of
10 items can be displayed on a Run screen. A Run screen configured for 5
items or less displays the items in the large font; a Run screen with 6-10
items displays the items in the small font.
●
In the Main Menu, choose Service >
Run Screens >
Select a Run
Screen
>
Number of Items
.
SET NUMBER OF ITEMS:
CURRENTLY: 5
SET TO: 4 ?
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Item
number
menu selection is used to assign a selected parameter to a
position in the Run screen display list. Item 1 is displayed at the top of the
list. If “None” is selected, the associated item will not be displayed on the
Run screen and will skip to the next item.
The available data fields are organized into the following groups; Other
Measurements, Other Measurements 2, Analog Inputs, and Non-
Measurements. If the I/O expansion board is not present, Analog Inputs
will not be available.
●
In the Main Menu, choose Service >
Run Screens >
Select a Run
Screen
>
Item
number
.
SELECT RUN SCREEN ITEM:
>OTHER MEASUREMENTS
ANALOG INPUTS
NON-MEASUREMENTS
OTHER MEASUREMENTS 2
CTRST SSET INSTS STAT
The Other Measurements screen allows the user to select an item that will
be included in the Run screen, such as flow volume or pump pressure.
Refer to Item Number described previously.
●
In the Main Menu, choose Service >
Run Screens >
Select a Run
Screen
> Item
number
>
Other Measurements
.
Number of Items
Item
Number
Other Measurements