Calibration
Flow Calibration
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
8-13
Figure 8–6.
Dual Filter Holder in the Filter Exchange
Figure 8–7.
Filter Exchange Closed
15.
From the calibration screen scroll to Flow2 AND PRESS ENTER.
Begin the calibration the unit should begin to bring the flow up to the
first set point.
ALLOW THE FLOW TO STABILIZE FOR AT
LEAST 15 MINUTES. THIS WILL ACT AS THE FLOW
CONTROLLER WARM-UP PERIOD.
16.
Following the 15 minute warm-up take a stabilized flow reading and
enter the measured value for each point. Wait 2-3 minutes for each
point after the first for stabilized reading and measurement.
17.
At the end of the calibration the unit will display the calculated
intercept or offset and span values necessary to linearize the flow sensor
to the operational range.
18.
Calibration procedure complete perform the audit per the operating
manual procedure.