Troubleshooting
System Status
11-2
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Table 11–1.
Diagnostic Menu Selections
Menu Item
Description
Program
Versions
Shows the version numbers of the installed programs (read only).
Voltages
Displays the current diagnostic voltage readings (read only).
Analog Input
Readings
Displays the eight user-scaled analog readings (if the I/O expansion board
option is installed).
Analog Input
Voltages
Displays the eight raw analog voltage readings (if the I/O expansion board
option is installed).
Digital Inputs
Displays the state of the 16 digital inputs (read only).
Relay States
Displays the state of the ten digital outputs, eight solenoid outputs, and
allows toggling of the state to either On (1) or Off (0).
Test Analog
Outputs
Used to set each of the analog output channels to zero or full scale. Channel
choices include all analog outputs, six voltage channels, and six current
channels (if the I/O expansion board option is installed).
Contact
Information
Displays telephone numbers of the customer service support centers.
The System Status menu provides access to system troubleshooting
information including status codes and alarms. The Status Codes screen
displays a description of the currently active status conditions. Refer to
Table 9–5
on page
9-63
for a list of the status codes.
●
In the Main Menu, choose System Status >
Status Codes
.
When an alarm is detected, the status of that associated item will go from
OK to FAIL. The Alarms menu displays a list of items that are monitored
by the Sampler. Refer to “
Alarms
” on page
9-64
for additional alarms
information.
●
In the Main Menu, choose System Status >
Alarms
.
This troubleshooting guide is designed to help isolate and identify
instrument problems.
Table 11–2
provides general troubleshooting
information and indicates the checks that you should perform if you
experience an instrument problem.
For additional service assistance, see “
Service Locations
” on page
11-22
.
System Status
Alarms
Troubleshooting