Sampler Operation
Post-Sampling Verification and Data Retrieval
7-20
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Use the following procedure to display Sampler Data on the instrument
LCD display.
1.
To display filter data, in the Main Menu, scroll to View Records, press
, with the cursor at Filter, press
, to display the filter
records. Filter Record 1 is the oldest record. Filter Record 2 is the most
recent record of the 32 maximum filters.
2.
Press
or
to display the different filter types. Press
or
to display the individual records.
3.
Scroll up or down through the filter record data to locate the desired
filter record. Scroll right to locate the status conditions that may have
been present during the sample run. The status information is also
provided with the downloaded instrument data. Record data from the
Filter Data screen onto a sampling run log sheet if desired. If there were
any status codes other than “OK,” verify the validity of the sampling
run by pressing scrolling to view the specific Status Codes.
Check the sampling run status on the Main screen, and note if the alarm
indicator is present. If the alarm indicator is displayed, press the STATS
soft key to display any status errors that are currently set. The Status screen
displays current staus conditions, but does not display status conditions
from previous completed samples. Refer to Chapter 9 “
Firmware Screen
Descriptions
” for more information on the status codes.
Display Sampler
Data
Identifying Sampler
Status