Firmware Screen Descriptions
Audit and Calibration Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-61
LEAK CHECK:
>INTERNAL
EXTERNAL
CTRST SSET INSTS STAT
The Leak Check Internal screen is used to perform an internal leak check.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Leak
Checks >
Internal
.
LEAK CHECK INTERNAL:
MAKE SURE THAT A LEAK
CHECK DISK IS IN THE
SAMPLE FILTER LOCATION!!
PRESS
TO BEGIN TESTING
CTRST SSET INSTS STAT
LEAK CHECK INTERNAL:
FILTER PRES: 80
TARGET: 55
VALVES AND PUMP ON
WAITING FOR PRESSURE
PRESSURE NOT REACHED !!
CTRST SSET INSTS STAT
The Leak Check External screen is used to perform an external leak check.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Leak
Checks >
External
.
LEAK CHECK EXTERNAL:
MAKE SURE THAT FILTER IS
IN PLACE AND LEAK CHECK
ADAPTER IS CLOSED!!!
PRESS
TO BEGIN TESTING
CTRST SSET INSTS STAT
Leak Check Internal
Leak Check External