Firmware Screen Descriptions
Audit and Calibration Menu
9-60
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
From In the Main Menu, choose Instrument Setup >
Flow Calibration
Setup > Flow
>
FTS Constant A
. The Flow Constant A screen allows the
user to adjust the setpoint. The FTS Constant B screen is used in the same
manner as the FTS Constant A screen.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
> Cal Setup >
Flow Constant A/B
.
FLOW CONSTANT A:
CURRENTLY: 1.0
SET TO: 1.
0
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Calibrate Ambient RH screen contains the current and latest averaged
values of the ambient relative humidity (%), as measured by a sensor
located in the ventilated filter compartment.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Ambient RH
.
CALIBRATE AMBIENT %RH:
CURRENTLY: 080 %
SET TO: 08
0
%
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Leak Check menu provides access to the internal leak check
parameters and the external leak check parameters used in performing leak
checks.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Leak
Checks
.
Flow Constant A/B
Ambient RH
Leak Checks