AK Protocol
Commands List
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
E-3
Table E–1
lists the Sampler AK protocol commands. The interface will
respond to the command strings outlined below.
Table E–1.
Commands List
Command
Description
Page
AKID
Returns identification information
E-3
AREG
Ask Register Command. This allows the user to query the Sampler for
the current value of any system variable (Program Register Code).
E-4
ASTO
Ask Storage Command. This allows the user to download a specified
number of records from the internal data logger from the current
position of the AK storage pointer. The location of this storage pointer
may be defined by the SSTO command.
E-6
EREG
Enter Register Command. This allows the user to assign a new value to
any system variable. Great care must be exercised in using this
command, as the value of variables should be changed only when the
monitor is in the appropriate operating mode.
E-7
ESAV
Saves parameters to flash as set by the EREG command.
E-8
SFxx (01,
02)
Set Function xx Command. This allows the user to send commands such
as <RUN/STOP> to the unit. Each command is designated with a two-
digit code, xx.
E-9
SSTO
Set Storage Command. This allows the user to change the location of
the AK storage pointer in the internal data logger, and is used in
conjunction with the ASTO command described above. The AK storage
pointer is always located just following the last record transmitted
through the RS232 port via the AK Protocol. If the circular buffer
overwrites this location or if the ASTO or SSTO commands have not
been used, the AK storage pointer resides at the oldest record in the
internal data logger. The current record type affected by the SSTO
command is denoted by PRC register 31.
E-10