AK Protocol
Enter Register Command (EREG)
COM 2-WAY SETTINGS
RS-Para 1
52
ASCII code for the 1-digit Station Number (for example "4": 052). The Station Number is always 1 digit in length.
RS-Para 2
75048
ASCII code representation of the 2-digit Channel Number (for example: "K0": 075, 048). The Channel Number is
always 2 digits in length.
RS-Para 3
13010
Optional: Up to 3 ASCII codes can be added to response from instrument. In this case, <CR> and <LF> (ASCII
codes 013 and 010) are appended to the response. Enter 0 if nothing is to be appended.
RS-Para 4
0
Not used.
Transmission to Instrument
Response from Instrument
Byte
Example
Description
B
No Err
Error
Description
1
<STX>
ASCII code 002.
1
<STX>
<STX>
ASCII code 002.
2
4
1-digit Station Number, RS-Para 1.
2
4
4
1-digit Station Number, RS-Para 1.
3
E
Enter Register command. Enter
the Program Register Code in
bytes 21 to 23 below, and the new
value of the variable in bytes
3
E
E
4-digit Enter Register command.
4
R
4
R
R
5
E
5
E
E
6
G
6
G
G
7
<space>
Space.
7
<space>
<space>
Space.
8
K
2-digit Channel Number, as
defined by RS-Para 2.
8
0
0
Number of current status conditions.
9
0
9
<space>
<space>
Space.
10
<space>
Space.
10
3
S
Program Register Code of the
variable whose value was entered.
The PRC may be 1 to 3 characters
long, and is not right-filled in the
response.
11
3
Program Register Code of the
variable whose value is being
entered. The PRC may be up to 3
digits long. Do not right-fill if the
PRC is less than 3 characters
long.
11
1
E
12
1
12
<ETX>
13
13
<ETX>
<CR>
ASCII code 003
14
<space>
Space.
14
<CR>
<LF>
Up to 3 digits appended to the end of
the response transmission, according
to the entry for RS-Para 3.
15
1
New value to be entered for
variable referenced by Program
Register Code in bytes 11 to 13
above.
NOTE: The value entered may be
of varying length.
15
<LF>
16
16
17
17
18
18
19
<ETX>
ASCII code 003.