-D Instruction Manual
4-1
Chapter 4
Partisol 2000
i
Inlet Conversion
The 2000
i
Sampler can be configured with a variety of inlet systems. This
section describes the components of the PM
2.5
VSCC inlet and WINS inlet
Sampler configurations, the PM
10
Sampler configuration, and the
procedures for converting the Sampler from a PM
2.5
to a PM
10
Sampler or
from a PM
10
to a PM
2.5
Sampler. Because the Partisol 2000
i
-D comes
equipped with a PM
10
inlet and virtual impactor for PM
2.5
and PM
coarse
sampling, this section does not apply to the Partisol 2000
i
-D.
For details, see the following topics:
●
“
Partisol 2000i PM2.5 VSCC Components
” on page
4-1
●
“
Partisol 2000i PM2.5 WINS Components
” on page
4-2
●
“
Partisol 2000i PM10 Components
” on page
4-3
●
“
Converting a Partisol 2000i from a PM2.5 Sampler to a PM10
Sampler
” on page
4-4
●
“
Converting a Partisol 2000i from a PM10 Sampler to a PM2.5
Sampler
” on page
4-7
The following components comprise the VSCC inlet configuration as
shown in
Figure 4–1
and
Figure 4–2
.
Figure 4–1.
PM
2.5
VSCC Cyclone, VSCC Adapter, Sample Tube, and U.S. EPA
PM
10
Inlet
Partisol 2000
i
PM
2.5
VSCC
Components
PM
10
Inlet
VSCC
Sample Tube
VSCC Adapter