Communications Using iPort or RPComm
Using RPComm
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
D-9
The Communication Port Setup screen will appear (
Figure D–7
). For
the 2000, the settings should be: baud rate = 9600, data bits = 8, parity
= none, stop bits = 1, flow control = none.
Note
If you are using multiple connections to different instrument types
and one instrument being used is a TEOM Series 1400 Ambient
Particulate Monitor, the flow control setting should be RTS.
▲
Figure D–7.
Communication Port Setup Screen
9.
If desired, additional connections can be defined using the above
procedure. These new connections can connect various instrument
types. When all new connections have been defined, they can be saved
in a connection group. From the Communicator Main screen (
Figure
D–9
), select the “File” pull down menu and choose “Save Group.” The
next time RPComm is executed, the connection group can be restored
by choosing “Open Group” from the “File” pull down menu of the
Communicator Main screen. A list of saved connection groups will be
displayed in the Save Connection Group screen (
Figure D–8
). Choose
the desired group and select “OK.”