Calibration
Filter Compartment Temperature Calibration
8-4
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Use the following procedure to calibrate the filter compartment
temperature.
1.
Press
to return to the Calibration menu, scroll to F Comp Temp
and press
.
2.
Determine the current temperature (°C) at the location of the filter
compartment probe in the ventilated filter compartment of the Partisol
Sampler using an external thermometer, [°C = 5/9 x (°F - 32)].
3.
At the F Comp Temp screen, use the pushbuttons to enter the
reference value in °C [°C = 5/9 x (°F - 32)] and press
to save the
value. The Sampler automatically adjusts the corresponding offset
based on this input.
Use the following procedure to perform a filter temperature calibration.
1.
Press
to return to the Calibration menu, scroll to Filter Temp
and press
.
If the calibration is done at a temperature lower or higher than ambient
temperature, the reference thermometer and Sampler probe should be
banded together and immersed to the same depth in an insulated
constant temperature bath or block. Reference and Sampler probes
should be equilibrated at temperature for at least 5 minutes before the
temperature is measured.
2.
Determine the current temperature (°C) at the location of the sample
filter in the Sampler using an external thermometer, [°C = 5/9 x (F –
32)].
3.
At the Calibrate Filter Temperature screen, enter the reference
temperature reading in °C [°C = 5/9 x (°F - 32)] and press
to
save the value. The Sampler automatically adjusts the corresponding
offset based upon this input.
Filter Compartment
Temperature
Calibration
Filter
Temperature
Calibration