Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
10-1
Chapter 10
Preventive Maintenance
The Technical Support Department at Thermo Fisher Scientific can be
consulted in the event of problems. In any correspondence with the factory,
please note both the serial number and program number of the instrument.
This chapter provides the following troubleshooting and service support
information:
●
“
Safety Precautions
” on page
10-1
●
“
Routine Maintenance Items
” on page
10-1
●
“
Inlet Maintenance
” on page
10-2
●
“
Cyclone Maintenance
” on page
10-6
●
“
WINS PM2.5 Impactor Maintenance
” on page
10-9
●
“
Virtual Impactor Maintenance Partisol 2000i-D
” on page
10-12
●
“
V-Seals Cleaning and Replacement
” on page
10-15
●
“
Service Locations
” on page
10-19
Note
The routine maintenance consists of the following procedures
performed at the indicated intervals.
▲
Read the safety precautions in the “About This Manual” section at the
front of this manual before performing any actions listed in this chapter.
The routine maintenance of the Partisol Samplers consists of the following
procedures performed at the indicated intervals (U.S. EPA 2.12
Quality
Handbook
, Section 9):
Filter Cassettes
- Inspect filter cassettes for contamination or damage after
every use. Discard any damaged cassettes. Wipe with a clean dry cloth as
required. Additional filter cassettes can be ordered from Thermo Fisher
Scientific (59-005923-0001). Do not expose the filter cassettes to
temperatures above 50 °C.
Safety
Precautions
Routine
Maintenance Items