Preventive Maintenance
Virtual Impactor Maintenance Partisol 2000i-D
10-12
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
12.
Screw the top of the WINS impactor back onto the bottom section.
Ensure that the WINS impactor remains in the upright orientation so
that the oil in the impactor assembly does not spill.
13.
Multiple WINS impactors can be prepared in this manner at one time
and kept for later use.
Use the following procedure to reinstall the WINS Impactor:
1.
Attach the adapter to the lower section of the WINS impactor.
2.
Push the WINS impactor (with its adapter attached) upward into the
WINS impactor mounting sleeve.
3.
Pass the rollers of the filter exchange mechanism through the slots in
the left-hand and right-hand guides.
4.
Allow the filter exchange mechanism to open fully.
5.
Insert the filter cassette carrier (with a filter cassette installed) into the
filter platform by matching the slot and hole in the filter cassette carrier
with the appropriate hardware.
6.
Push the handle of the filter exchange mechanism toward the back to
install the filter cassette.
7.
Perform a system leak check.
8.
If in the Audit mode, at the Main Menu, select Audit and Calibration
> Audit Mode, and press
to turn off the Audit mode.
Use the following procedures to remove, clean, and reinstall the virtual
impactor that is used in the Partisol 2000
i
-D Sampler.
Virtual impactor maintenance and performance verification can be
performed while in the Stop, Wait, or Sampling modes. If the Sampler is in
the Stop Mode, press
and select Audit and Calibration > Audit
Mode, and press
to change to the Audit mode.
WINS Impactor
Reinstallation
Virtual Impactor
Maintenance
Partisol 2000
i
-D