Firmware Screen Descriptions
Instrument Setup Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-21
SAMPLE METHOD:
CURRENTLY: BASIC
SET TO: BASIC
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Flow Rate screen is used to change the sample flow rate (flow rates for
the 2000
i
-D) for each sample.
2000
i
- Typically, flow rate is set to 1.67 L/min.
2000
i
-D
- Typically, flow rate is set to 15.0 L/min and flow2 rate is set to
1.67 L/min.
●
In the Main Menu, choose Instrument Setup >
Default Sample Setup >
Flow Rate/Flow2 Rate
.
SET FLOW RATE:
CURRENTLY: 15.00 L/min
SET TO: 15.0
0
L/min
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Filter Type screen allows the user to identify the filter type.
The Filter2 Type screen (2000
i
-D only) is used to identify the filter type
for the 2000
i
-D.
●
In the Main Menu, choose Instrument Setup >
Default Sample Setup >
Filter Type/Filter2 Type
.
FILTER TYPE:
CURRENTLY: A
SET TO: A
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Default Sample Times menu is used to set the default start time (start
time2 2000
i
-D only), duration/duration2 (2000
i
-D only).
Flow Rate/Flow2 Rate
Filter Type/Filter2 Type
Default Sample
Times