C-Link Protocol Commands
Communications Configuration
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
B-27
Note
The instrument will always respond to the command “mode” with
the status of the password lock as “mode local” or “mode remote”,
regardless of the above setting.
▲
The following example sets the instrument to accept the “set mode local”
and “set mode remote” commands.
Send:
set allow mode cmd 1
Receive:
set allow mode cmd 1 ok
Table B–6.
Allow Mode Command Values
Value
Allow Mode Command
0 Ignore
(default)
1 Accept
power up mode
This command reports the current power up mode setting to either 0 =
local/unlocked or 1 = remote/locked, according to
Table B–7
. The default
value is 0; power up in local/unlocked mode. The following example shows
that the instrument is configured to power up in the remote/locked mode.
Send:
power
up
mode
Receive:
power up mode 1
set power up mode
value
This command is used to configure the instrument to power up in the
local/unlocked mode (
value
= 0
)
or the remote/locked mode (
value
= 1),
according to
Table B–7
.
If the instrument is set to power up in the local/remote mode, the keypad
can be used to make changes via the front panel. If the instrument is set to
power up in the remote/locked mode, changes can not be made from the
front panel. The following example sets the instrument to power up in
remote/locked mode.
Send:
set
power
up
mode
1
Receive:
set power up mode 1 ok