Preventive Maintenance
V-Seals Cleaning and Replacement
10-18
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
vitual impactor to remove the virtual impactor to access the V-seal
(
Figure 10–14
).
Figure 10–14.
V-Seal on Inlet Adapter Tube on 2000
i
-D
The in-line filters (32-002643) are located in the lower section of the
Sampler, behind the sample pump, and protect the Sampler’s flow
controllers from particulate matter in the flow controller lines. The 2000
i
contains one inline filter; the 2000
i
-D contains two inline filters.
Use the following procedure to exchange the particle trap filters.
1.
Turn Off the Sampler.
2.
Locate the in-line filters behind the sample pump.
3.
Remove and replace the filters.
4.
Turn on the Sampler and perform a system leak check.
In-Line Filters
Replacement