Communications Using iPort or RPComm
Using RPComm
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
D-41
37.
The remaining portions of the screen should have been set while
establishing the modem connection. Do not modify these settings.
38.
Select “OK” to exit the AK Protocol Setup screen. This will display the
Connection Type screen.
39.
Choose “OK” from the Connection Type screen to finish the
connection setup.
40.
Connect to one of the instruments by highlighting the connection
name on the Connect List screen (
Figure D–20
) and select the
Connection icon on the tool bar. The connection to the instrument
will be initiated and the proper command codes sent. When
communication is established, the instrument’s main screen will be
displayed.
Note
If the connection is successful, the unit’s serial number will be
displayed at the top of the screen. Ensure that the correct serial number is
displayed to verify that the proper command codes were sent and that the
serial port trigger is functioning properly. If the connection is not successful
or if there is no instrument attached, then the serial number area will be
blank or will display “99999.” If the modem loses its connection, wait at
least 1 minute before trying to establish another connection. This allows
the modem to reset itself.
▲
41.
RPComm can now be used as described in “
RPComm
” on page
D-1
.
42.
To connect to the other instrument through the serial switching device,
the current connection must be terminated. Enter the proper command
codes for the desired instrument as described above and initiate the
modem connection. If the modem loses its connection, wait at least 1
minute before trying to establish another connection. This allows the
modem to reset itself.