Firmware Screen Descriptions
Diagnostics
9-92
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
CTRST SSET INSTS STAT
The Analog Input Readings screen (read only) displays the eight user-scaled
analog input readings (if the I/O expansion board option is installed).
●
In the Main Menu, choose Diagnostics
> Analog Input Readings
.
ANALOG INPUT READINGS:
>IN1 -0.57 V
IN2 -0.57 V
IN3 -0.57 V
IN4 -0.57 V
IN5 -0.57 V
CTRST SSET INSTS STAT
The Analog Input Voltages screen (read only) displays the eight raw analog
input voltage readings (if the I/O expansion board option is installed).
●
In the Main Menu, choose Diagnostics
> Analog Input Voltages
.
ANALOG INPUT VOLTAGES:
>ANALOG IN 1 6.24 V
ANALOG IN 2 4.28 V
ANALOG IN 3 0.00 V
ANALOG IN 4 0.00 V
ANALOG IN 5 0.00 V
CTRST SSET INSTS STAT
ANALOG IN 8 0.00 V
The Digital Inputs screen (read only) displays the state of the 16 digital
inputs. If nothing is connected to an input, the internal electronics will pull
the voltage up to high and the input will read (1). If a device is connected
to an input, the user configures whether the relay is normally open or
normally closed. This will dictate whether the input is brought to high (1)
or to ground (0) to call for an action.
●
In the Main Menu, choose Diagnostics
> Digital Inputs
.
Analog Input Readings
Analog Input Voltages
Digital Inputs