Firmware Screen Descriptions
Audit and Calibration Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-59
FLOW CAL SETUP:
>SETPOINT 1
SETPOINT 2
SETPOINT 3
SETPOINT 4
SETPOINT 5
CTRST SSET INSTS STAT
NO OF CAL POINTS 1.00
RESET FLOW CAL 0.00
FTS CONSTANT A
FTS CONSTANT B
The Setup Setpoint screen is used to adjust the setpoint.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
> Cal Setup >
Setpoint 1-5
.
SETUP SETPOINT:
CURRENTLY 12.000 l/min
SET TO: 12.00
0
l/min
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Set Number of Points screen is used to select the number of
calibration points.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
> Cal Setup >
No of Cal Points
.
SET NUMBER OF POINTS:
CURRENTLY 5
SET TO: 5
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
Reset Flow Cal menu item is used to reset the flow calibration to the
default value.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
> Cal Setup >
Reset Flow Cal
.
Setpoint 1-5
Number of Cal Points
Reset Flow Cal