C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
B-17
stream per
This command reports the current time interval in seconds for streaming
data. The following example reports the streaming period is set to 10
seconds.
Send:
stream
per
Receive: stream
per
10
set stream per
number value
number value
= | 1 | 2 | 5 | 10 | 20 | 30 | 60 | 90 |120 | 180 | 240 | 300 |
This command sets the time interval between two consecutive streaming
data strings to
number value
in seconds. The following example sets the
number value to 10 seconds.
Send:
set stream per 10
Receive:
set stream per 10 ok
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to
Table B–3
. The following example
reports that the streaming data shall not include a time stamp.
Send:
stream
time
Receive:
stream time 0
set stream time
value
This command enables
value
, where
value
is to attach or disable time stamp
to streaming data string, according to
Table B–3
. The following example
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–3.
Stream Time Values
Value
Stream Time
0
Disables time stamp to streaming data string
1
Attaches time stamp to streaming data string
frec
This command retrieves the most recent filter record. If there is no data,
the command returns “NO DATA”.
Send:
frec
Receive: frec
NO
DATA