The Spectrum Application (RF Measurements)
R&S
®
FPL1000
292
User Manual 1178.3370.02 ─ 03
●
Start and stop frequency
●
RBW
●
VBW
●
"Sweep Time"
●
"Sweep Points"
●
Reference level
●
Attenuator settings
●
Preamplifier settings
●
Transducer settings
●
Limit values
Via the sweep list, you define the ranges and their settings. For details on settings,
refer to
For details on defining the limits (masks), see
"Limit Lines in SEM Measurements"
For details on defining the limits (masks), see the base unit description "Working with
Lines in SEM".
Range definition
After a preset, the sweep list contains a set of default ranges and parameters. For
each range, you can change the parameters listed above. You can insert or delete
ranges.
The changes of the sweep list are only kept until you load another parameter set (by
pressing PRESET or by loading an XML file). If you want a parameter set to be availa-
ble permanently, create an XML file for this configuration (for details refer to
save a user-defined SEM settings file"
If you load one of the provided XML files, the sweep list contains ranges and parame-
ters according to the selected standard.
Reference range
The range containing the center frequency is defined as the reference range for all
other ranges in the sweep list. All range limits are defined in relation to the reference
range. The TX power used as a reference for all power level results in the result sum-
mary is also calculated for this reference range. You can define whether the power
used for reference is the peak power level or the integrated power of the reference
range. In the "Sweep List" , the reference range is highlighted in blue and cannot be
deleted.
Rules
The following rules apply to ranges:
●
The minimum span of a range is 20 Hz.
●
The individual ranges must not overlap (but can have gaps).
●
The maximum number of ranges is 30.
Measurements and Results