Remote Commands
R&S
®
FPL1000
678
User Manual 1178.3370.02 ─ 03
Parameters:
<State>
ON | 1
Includes the reference level offset in the results.
OFF | 0
Ignores the reference level offset.
*RST:
1
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Manual operation:
See
[SENSe:]PMETer<p>[:STATe]
<State>
This command turns a power sensor on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
PMET1 ON
Switches the power sensor measurements on.
Manual operation:
See
See
[SENSe:]PMETer<p>:UPDate[:STATe]
<State>
This command turns continuous update of power sensor measurements on and off.
If on, the results are update even if a single sweep is complete.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
PMET1:UPD ON
The data from power sensor 1 is updated continuously.
Manual operation:
See
UNIT<n>:PMETer<p>:POWer
<Unit>
This command selects the unit for absolute power sensor measurements.
Setting Basic Spectrum RF Measurement Parameters