
Remote Commands
R&S
®
FPL1000
594
User Manual 1178.3370.02 ─ 03
Parameters:
<Frequency>
Numeric value.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Manual operation:
See
[SENSe:]LIST:RANGe<range>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
For more information see
Chapter 7.2.7, "Spurious Emissions Measurement"
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Manual operation:
See
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation
<Attenuation>
This command defines the input attenuation for a spurious emission measurement
range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Attenuation>
Numeric value.
Refer to the data sheet for the attenuation range.
*RST:
10 dB
Default unit: dB
Manual operation:
See
Configuring and Performing Measurements