Remote Commands
R&S
®
FPL1000
659
User Manual 1178.3370.02 ─ 03
If you set the attenuation manually, it is no longer coupled to the reference level, but
the reference level is coupled to the attenuation. Thus, if the current reference level is
not compatible with an attenuation that has been set manually, the command also
adjusts the reference level.
Parameters:
<Attenuation>
Range:
see data sheet
Increment: 5 dB (with optional electr. attenuator: 1 dB)
*RST:
10 dB (AUTO is set to ON)
Example:
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
Manual operation:
See
INPut:ATTenuation:AUTO
<State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S
FPL1000 determines the signal level for opti-
mal internal data processing and sets the required attenuation accordingly.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Manual operation:
See
INPut:EATT:AUTO
<State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
Parameters:
<State>
1 | 0 | ON | OFF
1 | ON
0 | OFF
*RST:
1
Example:
INP:EATT:AUTO OFF
8.8.3.3
Configuring a Preamplifier
Setting Basic Spectrum RF Measurement Parameters