The Spectrum Application (RF Measurements)
R&S
®
FPL1000
373
User Manual 1178.3370.02 ─ 03
4. Disconnect all signals sending input to the power sensor and press ENTER to con-
tinue.
5. Wait until zeroing is complete.
A corresponding message is displayed.
How to Configure a Power Sensor as an External (PSE) Trigger
The following step-by-step instructions demonstrate how to configure a power sensor
to be used as an external power sensor trigger.
To configure a power sensor as an external power sensor (PSE) trigger
1. Connect a compatible power sensor to the "Power Sensor" interface on the front
panel of the R&S
FPL1000. (For details on supported sensors see
Sensor as an External Power Trigger"
2. Set up the power sensor as described in
"How to Set Up a Power Sensor"
3. In the "Power Sensor" tab of the "Input" dialog box, select the "External Power Trig-
ger" option.
4. Enter the power level at which a trigger signal is to be generated ( "External Trigger
Level" ) and the other trigger settings for the power sensor trigger.
5. Press the TRIG key and then select "Trigger/ Gate Config" .
6. In the "Trigger And Gate" dialog box, select "Signal Source" = "PSE" .
The R&S
FPL1000 is configured to trigger when the defined conditions for the
power sensor occur. Power measurement results are provided as usual.
7.3.3 Output Settings
Access
: "Overview" > "Output"
The R&S
FPL1000 can provide signals to different output connectors.
These connectors are only available if the R&S
FPL1-B5 option is installed.
For details on connectors, refer to the R&S
FPL1000 Getting Started manual, "Front /
Rear Panel View" chapters.
Receiving Data Input and Providing Data Output