The Spectrum Application (RF Measurements)
R&S
®
FPL1000
449
User Manual 1178.3370.02 ─ 03
Band power markers are only available for standard frequency measurements (not
zero span) in the Spectrum application.
For the I/Q Analyzer application, band power markers are only available for Spectrum
displays.
The entire band must lie within the display. If it is moved out of the display, the result
cannot be calculated (indicated by "-
-
-" as the "Function Result" ). However, the width
of the band is maintained so that the band power can be calculated again when it
returns to the display.
All markers can be defined as band power markers, each with a different span. When a
band power marker is activated, if no marker is active yet, marker 1 is activated. Other-
wise, the currently active marker is used as a band power marker (all other marker
functions for this marker are deactivated).
If the detector mode for the marker trace is set to "Auto" , the RMS detector is used.
The individual marker settings correspond to those defined in the "Marker" dialog box
(see
Chapter 7.9.2.1, "Individual Marker Setup"
on page 425). Any settings to the
marker state or type changed in the "Marker Function" dialog box are also changed in
the "Marker" dialog box and vice versa.
Marker Usage