![Rohde & Schwarz FPL1000 Series Скачать руководство пользователя страница 677](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826677.webp)
Remote Commands
R&S
®
FPL1000
677
User Manual 1178.3370.02 ─ 03
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
<NumberReadings>
This command sets the number of power readings included in the averaging process of
power sensor measurements.
Extended averaging yields more stable results for power sensor measurements, espe-
cially for measurements on signals with a low power, because it minimizes the effects
of noise.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<NumberReadings> An average count of 0 or 1 performs one power reading.
Range:
0 to 256
Increment: binary steps (1, 2, 4, 8, ...)
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
PMET2:MTIM:AVER:COUN 8
Sets the number of readings to 8.
Manual operation:
See
" Average Count ( Number of Readings )"
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
<State>
This command turns averaging for power sensor measurements on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation:
See
[SENSe:]PMETer<p>:ROFFset[:STATe]
<State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Setting Basic Spectrum RF Measurement Parameters