2. To configure several traces to predefined display modes in one step, press the but-
ton for the required function:
● "Preset All Traces"
● "Set Trace Mode Max | Avg | Min"
● "Set Trace Mode Max | ClrWrite | Min"
Chapter 7.10.1.2, "Trace Settings"
3. For "Average" trace mode, define the number of sweeps to be averaged in the
"Count:" field.
4. If linear scaling is used, select the "Average Mode" : "Linear" .
5. To improve the trace stability, increase the number of "Sweep Points" or the
"Sweep Time" (in the "Sweep" settings).
All configured traces (not set to "Blank" ) are displayed after the next sweep.
How to Copy Traces
1. A trace copy function is provided in a separate tab of the "Traces" dialog box. To
display this tab do one of the following:
● Select the TRACE key and then the "Trace Copy" softkey.
● Select "Analysis" from the "Overview" , then select the "Trace Copy" tab.
2. Select the "Source" trace to be copied.
3. Select the "Copy to Trace" button for the trace to which the settings are to be
applied.
The settings from the source trace are applied to the destination trace. The newly
configured trace (if not set to "Blank" ) is displayed after the next sweep.
7.10.1.4
How to Copy Traces
You can copy the trace settings from one trace to another in the "Copy Trace" tab of
the "Traces" dialog box.
► Select the "Source" trace and then the button for the "Copy To" trace.
Remote command:
7.10.2 Spectrograms
7.10.2.1
Working with Spectrograms
In addition to the standard "level versus frequency" or "level versus time" traces, the
R&S
FPL1000 also provides a spectrogram display of the measured data.
Trace Configuration