The Spectrum Application (RF Measurements)
R&S
®
FPL1000
419
User Manual 1178.3370.02 ─ 03
Some settings can be adjusted by the R&S
FPL1000 automatically according to the
current measurement settings. In order to do so, a measurement is performed. The
duration of this measurement can be defined automatically or manually.
Adjusting settings automatically during triggered measurements
When you select an auto adjust function a measurement is performed to determine the
optimal settings. If you select an auto adjust function for a triggered measurement, you
are asked how the R&S
FPL1000 should behave:
●
(default:) The measurement for adjustment waits for the next trigger
●
The measurement for adjustment is performed without waiting for a trigger.
The trigger source is temporarily set to "Free Run" . After the measurement is com-
pleted, the original trigger source is restored. The trigger level is adjusted as fol-
lows:
–
For IF Power and RF Power triggers:
Trigger Level = Reference Level - 15 dB
–
For Video trigger:
Trigger Level = 85 %
Remote command:
[SENSe:]ADJust:CONFigure:TRIGger
Adjusting all Determinable Settings Automatically ( Auto All )
.................................... 419
Adjusting the Center Frequency Automatically ( Auto Frequency )
............................ 420
Setting the Reference Level Automatically ( Auto Level )
........................................... 420
Resetting the Automatic Measurement Time ( Meastime Auto )
.................................420
Changing the Automatic Measurement Time ( Meastime Manual )
............................ 420
.............................................................................................. 420
.............................................................................................. 421
Adjusting all Determinable Settings Automatically ( Auto All )
Activates all automatic adjustment functions for the current measurement settings.
This includes:
●
Adjusting Settings Automatically