The Spectrum Application (RF Measurements)
R&S
®
FPL1000
334
User Manual 1178.3370.02 ─ 03
Remote command:
[SENSe:]LIST:RANGe<range>:RLEVel
RF Attenuation Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see
Chapter 7.5.1.2, "RF Attenuation"
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation:AUTO
RF Attenuation
Sets the attenuation value for that range.
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:GAIN:STATe
Sweep Points
Sets the number of sweep points for the specified range.
For details on sweep points, see
Chapter 7.6.1.8, "How Much Data is Measured:
Remote command:
[SENSe:]LIST:RANGe<range>:POINts[:VALue]
Stop After Sweep
This function is not available for the R&S
FPL1000.
Transducer
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
●
The transducer overlaps or equals the span of the range.
●
The x-axis is linear.
●
The unit is dB.
For details on transducers, see
Chapter 6.3.6.1, "Basics on Transducer Factors"
Remote command:
[SENSe:]LIST:RANGe<range>:TRANsducer
Limit Check
Activates or deactivates the limit check for all ranges.
For details on limit checks, see
"Limit Lines in Spurious Measurements"
"ABSOLUTE"
Signal is checked against absolute limit values
"NONE"
No limit check is performed.
Measurements and Results