Remote Commands
R&S
®
FPL1000
853
User Manual 1178.3370.02 ─ 03
End Try
End Sub
8.12 Deprecated Commands
The following commands are provided for compatibility to other signal analyzers only.
For new remote control programs use the specified alternative commands.
...................................................................................... 853
.........................................................................................853
.......................................................................................... 854
CALCulate<n>:LIMit<k>:TRACe
<TraceNumber>
This command links a limit line to one or more traces.
Note that this command is maintained for compatibility reasons only. Limit lines no lon-
ger need to be assigned to a trace explicitly. The trace to be checked can be defined
directly (as a suffix) in the new command to activate the limit check (see
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk
Suffix:
<n>
.
<k>
Parameters:
<TraceNumber>
1 to 6
*RST:
1
Example:
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
DISPlay[:WINDow<n>]:STATe
<State>
This command changes the display state of the selected measurement window.
Note that this command is maintained for compatibility reasons only. Use the
LAYout
commands for new remote control programs
(See
Chapter 8.7.1, "Working with Windows in the Display"
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
OFF | 0
The window is closed.
*RST:
0
Deprecated Commands