Remote Commands
R&S
®
FPL1000
797
User Manual 1178.3370.02 ─ 03
As a consequence, the results of a SCPI script using the
MMEMory:LOAD:STATe
com-
mand without a previous
MMEMory:SELect[:ITEM]
command may vary, depending
on previous actions in the GUI or in previous scripts, even if the script starts with the
*RST
command.
It is therefore recommended that you use the appropriate
MMEMory:SELect[:ITEM]
command before using
MMEMory:LOAD:STATe
.
Parameters:
1
<FileName>
String containing the path and name of the file to load.
The string may or may not include the file's extension.
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S
FPL1000.
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\Analyzer\user\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\Analyzer\user\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset instrument.
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\Analyzer\user\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the instrument.
(Switch the ON/OFF key off and on).
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\Analyzer\user\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Manual operation:
See
See
" Recall in New Channel / Recall in Current Channel "
MMEMory:STORe<n>:STATe
1,<FileName>
This command saves the current instrument configuration in a *.dfl file.
Suffix:
<n>
.
irrelevant
Parameters:
1
Managing Settings and Results