The Spectrum Application (RF Measurements)
R&S
®
FPL1000
239
User Manual 1178.3370.02 ─ 03
Figure 7-1: Sine wave signal with low S/N ratio
5. To suppress noise spikes, average the trace. In the "Traces" configuration dialog,
set the "Trace Mode" to "Average" (see
The traces of consecutive sweeps are averaged. To perform averaging, the
R&S
FPL1000 automatically switches on the sample detector. The RF signal,
therefore, can be more clearly distinguished from noise.
Figure 7-2: RF sine wave signal with low S/N ratio with an averaged trace
6. Instead of trace averaging, you can select a video filter that is narrower than the
resolution bandwidth. Set the trace mode back to "Clear/ Write" , then set the VBW
to
10
kHz
manually in the "Bandwidth" configuration dialog.
Measurements and Results