Remote Commands
R&S
®
FPL1000
757
User Manual 1178.3370.02 ─ 03
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
<DemodMode>
This command selects the demodulation mode at the marker position.
Suffix:
<n>
.
<m>
Parameters:
<DemodMode>
AM
AM demodulation
FM
FM demodulation
*RST:
AM
Example:
CALC:MARK:FUNC:DEM:SEL FM
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
<State>
This command turns the audio demodulator on and off when the measurement rea-
ches a marker position.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK3:FUNC:DEM ON
Switches on the demodulation for marker 3.
Manual operation:
See
[SENSe:]DEMod:SQUelch:LEVel
<Threshold>
This command defines the threshold for selective demodulation.
All signals below the threshold are not demodulated.
Parameters:
<Threshold>
Percentage of the display height.
Range:
0 to 100
*RST:
50
Setting Basic Spectrum RF Measurement Parameters