Remote Commands
R&S
®
FPL1000
559
User Manual 1178.3370.02 ─ 03
//Initiate a new measurement and waits until the sweep has finished.
INIT;*WAI
//---------------Limit Check--------------------
//Query the results of the limit check for the adjacent channels.
CALC:LIM:ACP:ACH:RES?
//Query the results of the limit check for the first alternate channels.
CALC:LIM:ACP:ALT1:RES?
//---------------Retrieving Results-------------
//Query the results for the ACLR measurement.
CALC:MARK:FUNC:POW:RES? ACP
8.6.4 Measuring the Carrier-to-Noise Ratio
The following commands are necessary to perform carrier-to-noise measurements.
●
CALC:MARK:FUNC:POW:SEL CN | CN0
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
●
[SENSe:]POWer:ACHannel:BANDwidth[:CHANnel<ch>]
●
Programming example: Measuring the carrier-to-noise ratio
This programming example demonstrates how to perform a Carrier-to-noise measure-
ment in a remote environment.
//----------Preparing the measurement------------
*RST
//Reset the instrument
FREQ:CENT 800MHz
//Sets the center frequency to the carrier frequency of 800 MHz.
CALC:MARK:FUNC:POW:SEL CN
//Activates carrier-to-noise ratio measurement.
POW:ACH:PRES CN
//Optimizes the instrument settings according to the channel bandwidth.
POW:ACH:PRES:RLEV
//Determines the ideal reference level for the measurement.
//--------------Performing the Measurement-----
INIT:CONT OFF
//Selects single sweep mode.
INIT;*WAI
//Initiates a new measurement and waits until the sweep has finished.
// Now turn off the carrier signal and repeat the measurement:
INIT;*WAI