Getting Started
R&S
®
FPL1000
21
User Manual 1178.3370.02 ─ 03
To ensure that the instrument software functions properly, certain rules must be
adhered to concerning the operating system.
Risk of rendering instrument unusable
The instrument is equipped with the Windows operating system. You can install addi-
tional software on the instrument, however, additional software can impair instrument
function. Thus, run only programs that Rohde
&
Schwarz has tested for compatibility
with the instrument software.
The drivers and programs used on the instrument under Windows are adapted to the
instrument. Only install update software released by Rohde
&
Schwarz to modify exist-
ing instrument software.
The following program packages have been tested:
●
R&S Power Viewer Plus - virtual power meter for displaying results of the power
sensor R&S NRP-Zxx (install only this component!)
●
Symantec Endpoint Security – virus-protection software
●
FileShredder - for reliable deletion of files on the hard disk
4.1.2.1
Virus Protection
Take appropriate steps to protect your instruments from infection. Use strong firewall
settings and scan any removable storage device used with a Rohde
&
Schwarz instru-
ment regularly. It is also recommended that you install anti-virus software on the instru-
ment. Rohde
&
Schwarz does NOT recommend running anti-virus software in the
background ("on- access" mode) on Windows-based instruments, due to potentially
degrading instrument performance. However, Rohde
&
Schwarz does recommend run-
ning it during non-critical hours.
For details and recommendations, see the following Rohde
&
Schwarz white paper:
●
1EF96: Malware Protection Windows 10
4.1.2.2
Service Packs and Updates
Microsoft regularly creates security updates and other patches to protect Windows-
based operating systems. These are released through the Microsoft Update website
and associated update server. Instruments using Windows, especially those that con-
nect to a network, should be updated regularly.
For details and recommendations, see the Rohde & Schwarz White Paper
4.1.2.3
Login
Windows requires that users identify themselves by entering a user name and pass-
word in a login window. By default, the R&S
FPL1000 provides two user accounts:
Preparing for Use