[SENSe:]ESPectrum<sb>:BWID
<Bandwidth>
This command defines the channel bandwidth of the reference range.
The bandwidth is available if the power reference is the channel power.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<Bandwidth>
minimum span ≤ value ≤ span of reference range
*RST:
3.84 MHz
Manual operation:
See
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa
<Alpha>
This command defines the roll-off factor for the RRC filter.
The RRC filter is available if the power reference is the channel power.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<Alpha>
Range:
0 to 1
*RST:
0.22
Manual operation:
See
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe]
<State>
This command turns the RRC filter in the reference range on and off.
The RRC filter is available if the power reference is the channel power.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
[SENSe:]ESPectrum<sb>:RRANge?
This command queries the reference range.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Configuring and Performing Measurements