Remote Commands
R&S
®
FPL1000
771
User Manual 1178.3370.02 ─ 03
Parameters:
<Type>
FREQuency | TIME
FREQuency
For limit lines that apply to a range of frequencies.
TIME
For limit lines that apply to a period of time.
*RST:
FREQuency
Example:
CALC:LIM:CONT:DOM FREQ
Select a limit line in the frequency domain.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:MODE
<Mode>
This command selects the horizontal limit line scaling.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST:
ABSolute
CALCulate<n>:LIMit<k>:CONTrol:OFFSet
<Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Offset>
Numeric value.
The unit depends on the scale of the x-axis.
*RST:
0
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters