Remote Commands
R&S
®
FPL1000
620
User Manual 1178.3370.02 ─ 03
//-----------Configuring the measurement ------------
*RST
//Reset the instrument
FREQ:CENT 100MHz
//Set center frequency
FREQ:SPAN 10KHz
// Set span
CALC:MARK:FUNC:MDEP ON
//Activate AM modulation depth measurement.
//--------------Performing the Measurement-----
INIT:CONT OFF
//Selects single sweep mode.
INIT:IMM
// Perform a single measurement
CALC:MARK:FUNC:MDEP:SEAR ONCE
//Initiates a search for signals in the current trace.
//---------------Retrieving Results-------------
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results.
//If the results are not accurate, change the position of the
//the temporary markers manually.
//----Changing the position of the temp markers-----
CALC:MARK:X 100MHZ
//Positions the reference marker on 100 MHz.
CALC:DELT2:X 5KHZ
//Positions delta marker 2 and 3 at a distance of 5 kHz to the reference marker.
CALC:DELT3:X 1KHZ
//Corrects the position of delta marker 3 by 1 kHz.
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results for the repositioned markers.
8.6.12 List Evaluations
A list evaluation is a multiple power measurement that measures the power at up to
200 frequencies. The measurement itself is a time domain measurement. Note that if
you set a span greater than 0, the R&S
FPL1000 aborts the list evaluation.
In case of a triggered measurement, a separate trigger event is required for each fre-
quency to initiate that measurement. Note that you have to make changes to the trig-
ger level in the time domain in order for it to take effect for the List Evaluation com-
mands.
Configuring and Performing Measurements