The Spectrum Application (RF Measurements)
R&S
®
FPL1000
344
User Manual 1178.3370.02 ─ 03
2. Set the center frequency to
1.8 GHz
.
3. Set the RBW to
100 kHz
.
4. Set the sweep time to
640
μ
s
.
5. Set the trigger source to "IF Power" .
6. Define a trigger offset of
-50
μ
s
.
7. Select the "Time Domain Power" measurement function from the "Select Measure-
ment" dialog box.
8. In the Time Domain Power configuration dialog box, set all four results to "On" .
9. Set the "Limit State" to "On" .
10. Define the left limit at
326
μ
s
and the right limit at
538
μ
s
.
This range corresponds to the useful part of the GSM burst.
The mean power of the useful part of the GSM burst is calculated to be -13 dBm.
7.2.9 Harmonic Distortion Measurement
The "Harmonic Distortion" measurement measures harmonics and their distortion,
including the total harmonic distortion.
........................................................................................345
...................................................................................345
..................................................................................347
Harmonic Distortion Configuration
........................................................................ 348
How to Determine the Harmonic Distortion
...........................................................350
Measurements and Results