Remote Commands
R&S
®
FPL1000
666
User Manual 1178.3370.02 ─ 03
"Incompleted sequential commands - blocked remote
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the TRIGGER INPUT connector.
IFPower
Second intermediate frequency
TIME
Time interval
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
" Using the power sensor as an external trigger "
on page 370
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
2.0 ms to 5000
Range:
2 ms to 5000 s
*RST:
1.0 s
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Manual operation:
See
8.8.4.2
Configuring Gated Measurements
................................................................................................ 667
...................................................................................667
.................................................................................... 667
Setting Basic Spectrum RF Measurement Parameters