[SENSe:]LIST:POWer:SET
<PeakPower>, <RMSPower>, <AVGPower>,
<TriggerSource>, <TriggerSlope>, <TriggerOffset>, <GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level
after
sending this command.
Parameters:
<PeakPower>
ON | OFF | 0 | 1
Turns peak power evaluation on and off.
*RST:
1
<RMSPower>
ON | OFF | 0 | 1
Turns RMS power evaluation on and off.
*RST:
0
<AVGPower>
ON | OFF | 0 | 1
Turns average power evaluation on and off.
*RST:
0
<TriggerSource>
Selects a trigger source.
<TriggerSlope>
NEGative | POSitive
Selects the trigger slop.
<TriggerOffset>
Defines the trigger delay.
Range:
negative measurement time to 30 s
*RST:
0
Default unit: s
<GateLength>
Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off.
To perform gated measurements, the trigger source must be dif-
ferent from
IMMediate
.
Range:
31.25 ns to 30 s
*RST:
0 s
[SENSe:]LIST:POWer:STATe
<State>
This command turns the List Evaluation off.
Parameters:
<State>
OFF | 0
*RST:
0